2 technical data, 2technical data – Nor-Cal CDG 100 OP Lit User Manual
Page 10

10
CDG100-OP-LIT (2010-1)
2
Technical Data
Measurement range
→ "Validity"
Accuracy
0.2% of reading
)
Temperature effect on zero
0.1 Torr (F.S.)
1 … 1000 Torr (F.S.)
0.0050% F.S./ °C
0.0025% F.S./ °C
Temperature effect on span
0.02% of reading / °C
Resolution
0.003% F.S.
Gas type dependence
none
Output signal analog
(measurement signal)
Measurement range
0 … +10 V
Voltage range
–5 … +10.24 V
(limited to +10.24 V)
Relationship voltage-pressure
linear
Output impedance
0
Ω (short-circuit proof)
Loaded impedance
>10 k
Ω
Response time
30 ms
130 ms
)
≥1 Torr (F.S.)
0.1 Torr (F.S.)
Identification
Resistance R
Ident
Voltage
13.2 k
Ω referenced to
supply common
≤5 V
1)
Non-linearity, hysteresis, repeatability in the calibrated range at 25 °C
ambient operating temperature without temperature effects after
operation of 2 h.
2)
Increase 10 … 90 % F.S.