Intended use, Functional principle, Trademark – Nor-Cal CDG 045 OP Lit User Manual
Page 4

4
CDG045-OP-LIT (2010-1)
The part number (P/N) can be taken from the product 
nameplate. 
If not indicated otherwise in the legends, the illustrations in this 
document correspond to Capacitance Diaphragm Gauges with 
the DN 16 ISO-KF vacuum connection. They apply to other 
vacuum connections by analogy. 
We reserve the right to make technical changes without prior 
notice. 
All dimensions in mm.
 
 
Intended Use
The temperature controlled Capacitance Diaphragm Gauges are 
intended for absolute pressure measurement of gases in their 
respective pressure ranges (
→  3).
The gauges can be operated in connection with a Nor-Cal 
Vacuum Gauge Controller or another appropriate controller. 
 
 
Functional Principle
A ceramic diaphragm is deflected by pressure. The deflection is 
measured capacitively and converted into an analog linear out-
put signal by the digital electronics. 
The output signal is independent of the gas type.
Very accurate pressure measurement is achieved by heating the 
sensor to a constant temperature of 45°C which results in a com-
pensation of changes in the ambient conditions and a reduced 
deposition of process products and by-products in process appli-
cations. 
 
 
Trademark
VCR
®
Swagelok Marketing Co.
 
 