beautypg.com

Connecting the electrical supply to the glove box – Labconco Precise HEPA-Filtered Glove Boxes 5220231 User Manual

Page 17

background image

Chapter 3: Getting Started

Product Service 1-800-522-7658

13

Connecting the Electrical Supply to the Glove Box

115V, 60 Hz; 100V, 50 Hz; or 100V, 60 Hz Models

Connect the power cord supplied to the IEC electrical supply plug on the
back of the glove box. 115V or 100V models are rated at 8 amps total.
The maximum circuit load for the interior electrical duplex is 5 amps.

230V, 50 Hz or 230V, 60 Hz Models

The same procedure applies for the 230V model except it is shipped without
a plug. Install the appropriate plug per local requirements. The 230V, 50 Hz
or 230V, 60 Hz models are both rated at 4 amps total. The interior
electrical duplex is rated at 2 amps.

Setting the Exhaust Volume with the Speed
Control Adjustment

For Precise HEPA-Filtered Glove Boxes, adjustment of the speed control
provides control of exhaust air volume and is located on the front panel. The
exhaust air volume should be adjusted at the low setting for weighing
operations. (Consult your Safety Officer for airflow recommendations
for your application).
Working at the lowest exhaust volume appropriate
for the application will give the quietest operation. To increase the rate of air
volume changes after working in the glove box and to ensure proper cleaning,
the speed control should be set at a higher (50 to 100) dial setting for 5 to 10
minutes as recommended by the operator’s predetermined protocol and the
chart below. The exhaust volume is increased by turning the speed control
counterclockwise.

Dial Setting

for Cleaning

Internal Air

Volume

Approximate

Air Volume

Changes per

Minute

Estimated

Total Exhaust

Volume

(CFM)

Minimum

Recommended

Times to

Clean Internal

Air Volume

(minutes)

50 4.7 61 10

100 7.7 100 5