Chapter 7: Modifying Your Freeze Dryer Display
Product Service: Domestic 1-800-522-7658, International 816-333-8811
Vacuum pump total hours to 0
Refrigeration to operate in Manual Mode
Temperature and Vacuum Offsets
If you do not want to restore all factory settings but want to
adjust offsets, press MENU and hold for approximately 10
seconds. The display will show
Press SELECT until the appropriate temperature offset is
shown. Each number represents 1°C.
Press VACUUM until the appropriate vacuum offset is
shown. Each number represents 0.001 mBar.
Press MENU to lock in the offsets and to return to the main
screen.
TEMPERATURE OFFSET: 0
VACUUM OFFSET:
0
This manual is related to the following products:
- FreeZone 1 Liter Benchtop Freeze Dry Systems 7740070 FreeZone 1 Liter Benchtop Freeze Dry Systems 7740061 FreeZone 1 Liter Benchtop Freeze Dry Systems 7740060 FreeZone 1 Liter Benchtop Freeze Dry Systems 7740041 FreeZone 1 Liter Benchtop Freeze Dry Systems 7740040 FreeZone 1 Liter Benchtop Freeze Dry Systems 7740031 FreeZone 1 Liter Benchtop Freeze Dry Systems 7740021 FreeZone 1 Liter Benchtop Freeze Dry Systems 7740020