Labconco FlaskScrubber Vantage Series Glassware Washers 4540031 User Manual
Page 59

Chapter 7: Troubleshooting
Product Service 1-800-522-7658
55
PROBLEM
CAUSE
CORRECTIVE ACTION
Low Temp Alarm
Heater inoperative
Check electrical connections.
Check voltage to heater.
Check resistance of heater. If open circuit,
replace heater.
Defective controller
Check output from PCB to heater.
Replace PCB. Contact Labconco Product Service.
Defective
temperature probe
Check electrical connections.
Contact Labconco Product Service. Replace if
necessary.
Defective high
temperature cut out
switch
Check resistance of safety cut out switch. If open
circuit at room temperature, replace switch.