3.1 PIN CONNECTIONS OF ALL UV VERSIONS

Pin connection scheme of the connector
[color of wire in yellow connection cable]:


1 - V+:

positive supply for signal circuit,

[red]

7 - 18 V, 1 W

2 - HEAT-GND: heater ground
[blue]

3 - UV-X-OUT:

UV-B or UV-E output, 0 - 3 V

[green]

not connected in UVS-A-T

4 - TEMP-OUT: internal temperature output

[yellow]

see table in Appendix II

5 - GNDA:

ground for signal outputs

[grey]

6 - HEAT V+:

positive supply voltage for heater

[brown]

7 - 18 V, 8 W


7 - UV-A-OUT:

UV-A output, 0 – 3 V

[white]

not

connected

in UVS-B-T or -E-T


8 - GNDA:

ground for signal circuit

[black]



Voltage drop over connection wires:

For correct operation of the sensor it is required that
the power supply and the connection cable have a total
resistance which does not exceed a critical value R

max.

This is to prevent voltage drop over the connection
wires that will reduce the supply voltage beyond the
lower operating limit of the radiometer electronics.