Specifications, 1 optical and electrical, 2 dimensions and weight – Kipp&Zonen CMA 6 Albedometers User Manual
Page 29: 1 optical and electrical 6.2 dimensions and weight
29
Instruction Manual - CMP/CMA series
.
Kipp & Zonen reserves the right to make changes to specifications and other product documentation without prior notice.
6.1 Optical and electrical
6.2 Dimensions and weight
CMP 3
CMP 6, CMP10, CMP 11, CMP 21 and CMP 22
CMA 6 and CMA 11
Weight without cable: 0.3 kg
Weight without cable: 0.6 kg
Weight without cable: 1.2 kg
6. Specifications
68 mm
Ø32 mm
Ø110 mm
Ø128 mm
Ø50 mm
Ø150 mm
464 mm
114 mm
Ø16 mm
68 mm
Ø50 mm
Ø150 mm
Ø150 mm
Specifications
CMP 3
CMP 22
CMP 21
CMP 6
Classification to ISO 9060:1990
Spectral range
(50% points)
Sensitivity
Impedance
Expected output range
(0 to 1500 W/m²)
Maximum operational irradiance
2000 W/m²
Response time
(63%)
Response time
(95%)
Zero offsets
(a)
thermal radiation
(at 200 W/m²)
(b)
temperature change
(5 K/h)
Non-stability
(change/year)
Second Class
300 to 2800nm
5 to 20µV/W/m²
20 to 200Ω
0 to 30mV
< 6s
< 18s
< 15 W/m²
< 5 W/m²
< 1%
Non-linearity
(100 to 1000 W/m²)
< 1.5%
Directional response
(up to 80° with 1000 W/m² beam)
< 20 W/m²
Spectral selectivity
(350 to 1500 nm)
< 3%
Temperature response
< 5%
(-10°C to +40°C)
Field of view
180°
Temperature sensor output
Operational temperature range
-40°C to +80°C
Humidity range
0 to 100% non-condensing
Tilt response
(0° to 90° at 1000 W/m²)
< 1%
Accuracy of bubble level
< 0.2°
Detector type
Thermopile
Storage temperature range
-40°C to +80°C
Ingress Protection (IP) rating
Recommended applications
67
Economical solution for
routine measurements in
weather stations,
field testing
4000 W/m²
Secondary Standard
285 to 2800nm
7 to 14µV/W/m²
10 to 100Ω
0 to 20mV
< 1.7s
< 5s
< 7 W/m²
< 2 W/m²
< 0.5%
< 0.2%
< 10 W/m²
< 3%
< 1%
(-20°C to +50°C)
180°
-40°C to +80°C
0 to 100% non-condensing
< 0.2%
< 0.1°
10K Thermistor
(optional Pt-100)
Thermopile
-40°C to +80°C
67
4000 W/m²
Secondary Standard
200 to 3600nm
7 to 14µV/W/m²
10 to 100Ω
0 to 20mV
< 1.7s
< 5s
< 3W/m²
< 1W/m²
< 0.5%
< 0.2%
< 5 W/m²
< 3%
< 0.5%
(-20°C to +50°C)
180°
-40°C to +80°C
0 to 100% non-condensing
< 0.2%
< 0.1°
10K Thermistor
(optional Pt-100)
Thermopile
-40°C to +80°C
67
2000 W/m²
First Class
285 to 2800nm
5 to 20µV/W/m²
20 to 200Ω
0 to 30mV
< 6s
< 18s
< 12 W/m²
< 4 W/m²
< 1%
< 1%
< 20 W/m²
< 3%
< 4%
(-10°C to +40°C)
180°
-40°C to +80°C
0 to 100% non-condensing
< 1%
< 0.1°
Thermopile
-40°C to +80°C
67
Good quality measurements
for hydrology networks,
greenhouse climate control
Note: The performance specifications quoted are worst-case and/or maximum values
Standard 10k Thermistor or optional Pt-100 temperature sensor with CMP 21 and CMP 22
Individual directional response and temperature dependence test data with CMP 21 and CMP 22
4000 W/m²
Secondary Standard
285 to 2800nm
7 to 14µV/W/m²
10 to 100Ω
0 to 20mV
< 1.7s
< 5s
< 7 W/m²
< 2 W/m²
< 0.5%
< 0.2%
< 10 W/m²
< 3%
< 1%
(-10°C to +40°C)
180°
-40°C to +80°C
0 to 100% non-condensing
< 0.2%
< 0.1°
Thermopile
-40°C to +80°C
67
Meteorological networks,
reference measurements in
extreme climates, polar
or arid
Scientific research
requiring the highest level
of measurement accuracy
and reliability
Meteorological networks,
PV panel and thermal
collector testing,
materials testing
CMP10 & CMP 11