LIST OF SPECIFICATIONS

Spectral range:

300 to 2800 nm, 50% points


Sensitivity: 7

µV/Wm

-2

(nominal)


Impedance:

500 to 2000

Ω


Response time:

18 s

(95% response)

< 8 s

(63% response)


Non-linearity:

Max. 3 % (0 - 2500 W/m

2

)


Directional error (at 80°
with a 1000 W/m² beam):

± 20 W/m²

Temperature dependence
of sensitivity:

3 % (-20

°C to +0 °C)

2 % (0

°C to +100 °C)

3 % (+100

°C to +150 °C)



Tilt error:

Max. 1% deviation when facing

downwards

Zero-offset due to
temperature changes:

Max. 4 W/m

2

offset for 5 K/h temp.

change.


Zero-offset due to FIR
(200 W/m

2

):

± 15 W/m

2


Operating temperature:

-40 °C to +150

°C


Field of view:

180° (2

π sr)


Max irradiance:

4000 W/m

2


Non-stability:

± 1% sensitivity change per year