Heath Consultants Remote Methane Leak Detector (RML) User Manual
Page 30
• Low battery
recharge if necessary
Excessive false detections
while scanning at longer
distances
• Scanning too fast
• Alarm detection
threshold set too low
• Scanning at the range
limit of the instrument
• Slow down the scanning
rate. Pause at the long
range and sweep
towards you
• Increase the Alarm
detection threshold
• Move in closer
Excessive false detections
while scanning at closer
distances
• Scanning too fast
• Alarm detection
threshold set too low
• Avoid making abrupt
motions while scanning
• Increase the Alarm
Detection Threshold
Excessive false detections
or loss of sensitivity
• Laser output not
optimized
• Alarm Detection
Threshold set too high
or low for conditions
• Perform Self-Test
procedure
• Check the Alarm
Detection Threshold
Error icon or Warning icon
on continuously
• Low battery
• Moisture condensation
on mirror due to rapid
change in temperature
• Internal component
failure
• Check battery level and
recharge if necessary
• Allow for the
temperature of the
instrument to stabilize
• Note error code and
contact HEATH
Battery indicator does not
show full charge after
charging
• Battery not fully
charged
• Battery level calibration
error
• Charge unit until solid
green light on charger is
on
• Run unit until it shuts
off, then fully recharge
with out interruption
Battery indicator bars are
flashing on and off
• Battery level calibration
error
• Run unit until it shuts
off, then fully recharge
with out interruption
Page 30 of 36
101515-0 RMLD MANUAL REV E