Field calibration procedure, Calibration procedure – Heath Consultants Gasurveyor 500 Series User Manual
Page 53

B-5
Field Calibration Procedure
1) Zero gas ranges before attempting calibration. See
previous section ZEROING THE INSTRUMENT
for details.
2) Make sure that the instrument pump is running and
the gas range selected is compatible with the
calibration gas.
Note: A single press of Button One
toggles the pump
Off / On.
3) Remove the cap from calibration gas cylinder.
Make sure that the regulator valve is in the fully
closed position (Off) then connect the gas regulator
to the gas cylinder (push down gently and tighten
clockwise, hand tight). See Figure B-5 for details.
Figure B-5 Connecting Gas
FIELD CALIBRATION