beautypg.com

Filter replacement – Heath Consultants Gasurveyor 500 Series User Manual

Page 34

background image

GASURVEYOR 3-500 USER HANDBOOK

24

Filter Replacement

Hydrophobic and cotton particulate filters in the probe
handle minimise the danger of water and dust ingress.

Caution: The instrument should never be switched on
without suitable filters installed.

If a blockage occurs the ‘SAMPLE FAULT’ indicator is
displayed. Check the sample line and probe handle for
blockage. Press Button One

to clear the ‘SAMPLE

FAULT’ message. Replace the filter(s) if the message does
not clear.
To replace the filter(s), proceed as follows:

1) Unscrew the probe handle assembly.

Hydrophobic Filter

Cotton Particulate Filter

Probe Handle Assembly

Figure 4.1 Filter Assembly