CHAPTER VI
Maintenance and Troubleshooting Information
Troubleshooting the Instrument
The advanced design of the DP-IR makes it one of the most reliable leak survey instruments 
available. However, should you experience problems with the instrument or suspect that the 
instrument is not operating properly, do not use the instrument for leak survey work until the 
problem is resolved.
Only a qualified DP-IR repair technician should attempt to repair or open the instrument. There 
are no user-serviceable components in the DP-IR that can be repaired or replaced other than the 
battery.
Most often a problem has a simple cause. The following table provides a list of common 
problems, causes and solutions. Should you have a problem not listed or the solution does not 
work, please contact 
HEATH for further assistance.
Symptom
Unit will not turn on
Loss of sensitivity
Excessive false alarming
ERROR icon and alarm on 
continuously
Probable Cause(s)
• Low battery
• Filters need replacing
• Alarm Level set too high for 
 conditions
• Zero drift
• Alarm Level set too low for 
 conditions
• Filters need replacing
• Low battery
• Internal component failure
Solution
• Recharge internal battery
• Use the External Power 
 Supply
• Check filters and replace if 
 necessary
• Zero the instrument in a gas 
 free environment
• Lower the Alarm Threshold
• Perform Self-Test procedure
• Check filters and replace if 
 necessary
• Zero the instrument
• Raise Alarm Level
• Perform Self-Test procedure
• Check battery level and 
 recharge if necessary
• Run Self-Test