Hanna Instruments HI 9146-04 User Manual

Page 4

background image

6

7

PROBE PREPARATION

All probes are shipped dry by Hanna Instru-
ments. To hydrate the probe and prepare it
for use, connect it to the meter and proceed
as follows.
1 Remove the red and black plas-

tic cap which is for shipping
purposes and can be discarded.

2 Wet the sensor by soaking the

bottom 2½ cm (1") of the probe
in electrolyte (HI 7041S) for 5
minutes.

3 Take a membrane

(HI 76407A sup-
plied with the
meter) and Make
sure that the rub-
ber O-ring sits
properly inside
the membrane
cap.

4 Rinse the mem-

brane with electro-
lyte while shaking
it gently. Refill
with clean electro-
lyte.

5 Gently tap the sides of the membrane cap

with your finger tip to ensure that no air
bubbles remain trapped. To avoid damag-
ing to the membrane, do not tap the mem-
brane directly on the bottom.

6 With the sensor facing down, screw the

cap clockwise. Some electrolyte will over-
flow.

When not in use, protect the mem-
brane by placing the protective cap
on it.

Range

0.00 to 45.00 mg/L O

2

0.0 to 300 %O

2

0.0 to 50.0 °C

Resolution

0.01 mg/L O

2

0.1 %O

2

0.1 °C

Accuracy

±1.5% Full Scale mg/L O

2

(@25°C/77°F)

±1.5% Full Scale %O

2

±0.5 °C

Typical EMC

±0.3 mg/L O

2

/ ±3.5 %O

2

Deviation

±0.5 °C

Calibration

Automatic in saturated air

Temperature

Automatic from 0 to 50°C

Compensation

(32 to 122°F)

Altitude

0 to 4 km

Compensation

100 m resolution

Salinity

0 to 80 g/L

Compensation

1 g/L resolution

Environment

0 to 50°C (32 to 122°F)

RH 100%

Power supply

• 4x1.5V AA batteries;

200 hours of continuous use;

auto-off after 4 hours.

• 12 Vdc adapter

Dimensions

196 x 80 x 60 mm

(7.7 x 3.1 x 2.4")

Weight

500 g (1.1 lb.)

SPECIFICATIONS

Shipping

cap

black

red

THEN SCREW

BACK ON

THIRD

TAP

SECOND

FILL

FIRST

INSERT

O’RING