Hanna Instruments HI 83215 User Manual

Page 33

background image

33

Nitrate LR

SPECIFICATIONS

Range

0.0 to 30.0 mg/L

Resolution

0.5 mg/L

Accuracy

±1.0 mg/L ±8% of reading at 25 °C

Typical EMC

±0.5 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the cadmium reduction method. The reaction between nitrate-nitrogen
and the reagent causes an amber tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93728-0

Powder reagent

1 packet

REAGENT SETS

HI 93728-01 Reagents for 100 tests
HI 93728-03 Reagents for 300 tests
For other accessories see page 50.

MEASUREMENT PROCEDURE

Note: for sample preparation follow the IRRIGATION WATER (LR) procedure at page 19.

• Select the

Nitrate LR method using the procedure

described in the

Method Selection section (see page 12).

• Using the pipette, fill the cuvette with 6 ml of sample,

up to half of its height, and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the cuvette and add the content of one packet of

HI 93728-0 reagent.

NITRATE LOW RANGE

6 mL