Hanna Instruments HI 83742 User Manual

Page 14

background image

14

COLOR DENSITY, WHITE WINE

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS
Range

0.000 to 1.000

Resolution

0.001

Precision

±0.010 @ 0.200

Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Direct reading.

REQUIRED REAGENTS

REQUIRED REAGENTS

REQUIRED REAGENTS

REQUIRED REAGENTS

REQUIRED REAGENTS
No reagents required. For other accessories see page 26.

MEASUREMENT

MEASUREMENT

MEASUREMENT

MEASUREMENT

MEASUREMENT

• Press “RANGE” to select the parameter code “P1” for

color density - white wine (see page 13).

• Fill an empty cuvet with deionized water and replace

the cap. This is the zero

zero

zero

zero

zero.

• Insert the zero cuvet into the holder and close the lid.

• Press ZERO and “----” will blink on the display.

SAMPLE PREPARATION

SAMPLE PREPARATION

SAMPLE PREPARATION

SAMPLE PREPARATION

SAMPLE PREPARATION

• Use the plastic pipette to fill a cuvet with white wine.

and replace the cap. This is the sample.

#1

(Zero)

#2

(Sample)

15

• After a few seconds the display will show “-0.0-”.

The meter is now zeroed and ready for measurement.

• Remove the cuvet from the instrument.

• Insert the sample (cuvet #2) into the holder and

close the lid.

#2

(Sample)

• Press READ/TIMER and “----” will blink during

measurement.

• The instrument directly displays the color density on

the Liquid Crystal Display.

INTERFERENCES
Suspended matter should be removed by centrifugation or prior filtration. Use an adeguate filter
material that does not adsorb color.