MEASUREMENT PROCEDURE

• Turn the meter on by pressing ON/OFF.

• When the LCD displays “724”, it is ready.

• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row, but it is recommended to take a new zero

reading for each sample and to use the same cuvet for zeroing and measurement.

• After the reading it is important to discard immediately the sample, otherwise the glass

might become permanently stained.

• All the reaction times reported in this manual are referred to 20°C (68°F). As a general rule

of thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a measurement follow the validation procedure to be

sure that the instrument is properly calibrated. If necessary, calibrate the instrument.

• Fill the cuvet with 10 mL of unreacted

sample, up to the mark, and replace the
cap.

• Place the cuvet into the holder and ensure

that the notch on the cap is positioned
securely into the groove.

• Press ZERO and “SIP” will blink on the

display.

• After a few seconds the display will show

“-0.0-”. The meter is now zeroed and
ready for measurement.

• Remove the cuvet.

13

For free chlorine measurement: add 3 drops
of HI 93701A-T DPD1 indicator and 3
drops of HI 93701B-T DPD1 buffer to an
empty cuvet.
Immediately add 10 mL of unreacted
sample. Replace the cap and shake gently.

Immediately reinsert the cuvet into the
holder and ensure that the notch on the
cap is positioned securely into the groove.

• Press READ and “SIP” will blink during

measurement.

• The instrument directly displays

concentration in mg/L of free chlorine on
the Liquid Crystal Display.

• Then,

for total chlorine measurement, remove

the cuvet and add to the reacted sample 3
drops of HI 93701C-T DPD3 solution.
Replace the cap and swirl gently.

• Reinsert the cuvet into the holder and

ensure that the notch on the cap is
positioned securely into the groove.

• Wait for 2’ and 30” then press READ and

“SIP” will blink during measurement.

• The instrument directly displays

concentration in mg/L of total chlorine on
the Liquid Crystal Display.