beautypg.com

Delphi DL-00183 User Manual

Page 35

background image

23

DL-00096 Rev. 7

Immunity test

Electrostatic

discharge (ESD)

IEC 61000-4-2

Electric fast

transient/burst

IEC 61000-4-4

Surge

IEC 61000-4-5

IEC 60601 test
level

± 6kV contact

± 8kV air

± 2kV for

power supply

lines

± 1kV for

input/output

lines
± 1kV

differential mode

± 2kV common

mode

Electromagnetic
environment/guidance

Floors should be wood,

concrete, or ceramic tile.

If floors are covered with

synthetic material, the

relative humidity should

be at least 30%.
Mains power quality

should be that of a typical

commercial or hospital

environment.

Mains power quality

should be that of a typical

commercial or hospital

environment.

Guidance and manufacturer’s declaration: electromagnetic immunity

The Portable Oxygen Concentrator is intended for use in the electromagnetic
environment specified below. The customer or the user of the concentrator should
assure that it is used in such an environment.

Compliance
level

± 15kV contact

± 8kV air

± 2kV for

power supply

lines

± 1kV for

input/output

lines
± 1kV

differential mode

± 2kV common

mode

DL-00096 rev 7.indd 23

6/1/2009 4:48:35 PM