Appendix 2: technical information – Delphi DL-00183 User Manual
Page 34
22
DL-00096 Rev. 7
Appendix 2: Technical Information
ELECTROMAGNETIC COMPATIBILITY (EMC) INFORMATION
Medical electrical equipment requires special precautions regarding
electromagnetic compatibility (EMC). Portable and mobile radio frequency (RF)
communications equipment can affect devices such as the Portable Oxygen
Concentrator. As such, the device should not be used adjacent to other
equipment. If this is not practical, then observe the device to make sure it is
operating properly at all times.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic environment/guidance
The Portable Oxygen Concentrator uses
RF energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any interference
in nearby electronic equipment.
The concentrator is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Guidance and manufacturer’s declaration: electromagnetic emissions
The Portable Oxygen Concentrator is intended for use in the electromagnetic
environment specified below. The customer or the user of the concentrator should
assure that it is used in such an environment.
DL-00096 rev 7.indd 22
6/1/2009 4:48:35 PM