Alarm indicators – Delphi DL-00183 User Manual
Page 28
16
DL-00096 Rev. 7
If the Portable Oxygen Concentrator detects an alarm condition, it will indi-
cate the alarm visually and audibly within 10 seconds. There are four levels
of alarms: critical high priority, high priority, medium priority, and low priority.
Each is indicated differently by the backlit display; green, yellow, and red
LEDs; and buzzer, as indicated below. In each case, the alarm message and
power status will override the current display.
Note: All alarm conditions and parameters are factory preset; conditions and parameters cannot be
changed or adjusted by the user.
Alarm Status
Critical high
priority
High priority
Medium priority
Low priority
Audible Tone
Ten beeps per burst, burst
repeats every 3 seconds
Ten beeps per burst, burst
repeats every 3 seconds
Three beeps per burst,
burst repeats every 8
seconds
Three beeps per burst,
burst repeats every
10 minutes
Visual Indicator
Solid red LED
(and device shuts
off automatically)
Flashing red LED
Flashing yellow LED
Solid yellow LED
Note: If two alarm conditions exist at the same time, the highest priority alarm is indicated. If two or
more alarm conditions of equal priority exist at the same time, the most recent one will be displayed.
Note: The most recent alarms indicated by the device are logged for reference by service personnel.
This log is maintained even if the device is powered down or if power is lost for any other reason.
Note: If the mute key is pressed prior to an alarm condition (for example, to mute the device in a
movie theater), critical high priority and high priority alarms will override the mute function; medium
and low priority alarms will be muted for eight hours from the time the key was pressed. Press the
mute key off to display the last highest priority alarm. Press the mute key on again to reset the eight-
hour timer.
ALARMS
When the Portable Oxygen Concentrator sounds an alarm, a corresponding
message will be displayed on the screen. Take appropriate action as directed
in the charts below.
Critical High Priority Alarms
Note: These alarms will disable the Portable Oxygen Concentrator immediately.
Alarm Message
Charge Battery
XX: Service!*
*Value: 01-20
Description
Battery needs charging.
Service required.
Action
Recharge the battery
pack by plugging in to
power supply. Ensure all
connections are made
securely.
Contact Customer
Service.
Alarm Indicators
Mute Time
20 minutes
20 minutes
8 hours
24 hours
DL-00096 rev 7.indd 16
6/1/2009 4:48:34 PM