5

Range

0.000 to 0.500 mg/L

Resolution

0.001 mg/L

Precision

±0.004 mg/L @ 0.200 mg/L

Typical EMC Deviation

±0.001 mg/L

Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Light Detector

Silicon Photocell

Method

Adaptation of the USEPA METHOD 330.5. The reaction between the
chlorine and the DPD reagent causes a pink tint in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in measurement mode;
after 1 hour of non-use in calibration mode.

Dimensions

180 x 83 x 46 mm (7.1 x 3.3 x 1.8")

Weight

290 g (10 oz.).

REQUIRED REAGENTS
Code

Description

Quantity/test

HI 95761-0

Total Chlorine LR Reagent

1 packet

SPECIFICATIONS

PRECISION AND ACCURACY

Precision is how closely repeated
measurements agree with each other.
Precision is usually expressed as standard

deviation (SD).
Accuracy is defined as the nearness of a
test result to the true value.
Although good precision suggests good

accuracy, precise results can be inaccurate.
The figure explains these definitions.
In a laboratory using a standard solution of

0.200 mg/L chlorine and a representative
lot of reagent, an operator obtained with a
single instrument a standard deviation of
0.004 mg/L.

16

CALIBRATION PROCEDURE
• Turn the meter on by pressing ON/OFF.
• When the LCD displays “761”, it is ready.

• Enter the calibration mode by holding CAL CHECK

for three seconds.

• The date of the last calibration appears (e.g.:

month “01”, day “08”). “F.CAL” means that
the factory calibration is selected.
Note: at this point it is possible to reset the

instrument to restore FACTORY CALIBRATION,
see instructions page 18.

• Place the CAL CHECK™ Standard Cuvet A into

the holder and ensure that the notch on the cap
is positioned securely into the groove.

• Press ZERO and “SIP” will blink on the display.

• After a few seconds the display will show

“-0.0-”. The meter is now zeroed and ready for

calibration.

• Remove the cuvet.

• Place the CAL CHECK™ Standard HI 95761-11

Cuvet B into the holder and ensure that the
notch on the cap is positioned securely into the

groove.

Note: it is possible to interrupt calibration procedure at any time by pressing ON/OFF.
Warning: do not validate or calibrate the instrument with standard solutions other than the Hanna
CAL CHECK™ Standards, otherwise erroneous results will be obtained.

or

CALIBRATION PROCEDURE