• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.
The blinking “ZERO” indicates that the
instrument needs to be zeroed first.

• Fill one cuvette with 10 mL of unreacted

sample, up to the mark.

• Place the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• After a few seconds, the display will

show “-0.0-”. The meter is now zeroed
and ready for measurement.

• Add 4 drops of the HI 93700A-0 (6

drops in case of seawater analysis).

• Replace the cap and swirl the solution.

• Add 4 drops of the HI 93700B-0 (10

drops in case of seawater analysis).

• Replace the cap and swirl the solution.

• Replace the cuvette into the holder and

ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER for three

seconds. The display will show the
countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.