• Remove the cuvette.

• Add the content of one packet of

HI 93708-0

reagent. Replace the cap

and shake gently.

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 10 minutes then
just press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the instrument

directly displays concentration in mg/L of
nitrite on the LCD.

• To convert the NO

2

-

ion concentration to

NO

2

-

-N concentration, divide the reading

by a factor of 3.29.

• To convert the NO

2

-

ion to sodium nitrite

NaNO

2

concentration, multiply the reading

by a factor of 1.5.

INTERFERENCE
HI 96708

will not recognize any glycol based samples.