VALIDATION PROCEDURE

Calcium above

700 mg/L

Chloride above

70000 mg/L

Iron above

5 mg/L

Magnesium above

100000 mg/L

Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:

Do not validate the instrument with any standard solutions other than the HANNA CAL

CHECK

TM

Standards, otherwise erroneous results will be obtained.

• Turn the meter on by pressing ON/OFF.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• Place the CAL CHECK

TM

Standard

HI 96709-11

Cuvette A into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

INTERFERENCES

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.
The blinking “ZERO” asks for instrument
zeroing.