• Alternatively, wait for 3 minutes then just

press READ/ /TIMER. In all cases, the
lamp, cuvette and detector icons will
appear on the display, depending on the
measurement phase.

• At the end of measurement, the instrument

directly displays concentration in mg/L of
phosphate on the LCD.

• To convert the reading to mg/L of

phosphorus pentoxide (P

2

O

5)

, multiply by

a factor of 0.747.

• To convert the reading to mg/L of

phosphorus (P) concentration, multiply
by a factor of 0.326.

INTERFERENCES:

Interference may be caused by:
Iron above 50 mg/L

Silica above 50 mg/L

Silicate above 10 mg/L

Copper above 10 mg/L
Hydrogen sulfide, arsenate, turbid sample and highly buffered samples also interfere.