• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.
The blinking “ZERO” indicates that the
instrument needs to be zeroed first.

• Fill a graduated beaker with 50 mL of

sample.

• Add 0.5 mL of HI 93719A-0 Magnesium

indicator solution, then swirl to mix.

• In the same beaker add 0.5 mL of HI

93719B-0

Alkali solution for

Magnesium, then swirl to mix.

• Fill two cuvettes with 10 mL of sample

each.

• Add 1 drop of HI 93719C-0 EDTA solution

to one cuvette, replace the cap and swirl
the solution. This is the blank.

• Add 1 drop of HI 93719D-0 EGTA solution

to the second cuvette, replace the cap
and swirl the solution. This is the sample.

• Place the blank into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• After a few seconds, the display will

show “-0.0-”. The meter is now zeroed
and ready for measurement.