beautypg.com

Hanna Instruments HI 96720 User Manual

Page 7

background image

12

13

• It is important that the sample does not contain any debris. This would

corrupt the reading.

• Each time the cuvette is used, the cap must be tightened to the same

degree.

• Whenever the cuvette is placed into the measurement cell, it must be

dry outside, and completely free of fingerprints, oil or dirt. Wipe it
thoroughly with HI 731318 or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles in the sample, causing higher readings. To obtain

accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette.

• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.
• It is possible to take multiple readings in a row, but it is recommended to take a new zero

reading for each sample and to use the same cuvette for zeroing and measurement.

• After the reading it is important to discard immediately the sample, otherwise the glass

might become permanently stained.

• All the reaction times reported in this manual are referred to 20°C (68°F). As a general rule

of thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a measurement follow the validation procedure to

be sure that the instrument is properly calibrated. If necessary, calibrate the instrument.

MEASUREMENT PROCEDURE

STARTUP

Prepare the instrument for measurement as follows:
• Unpack the instrument by removing the dust protection sleeve from the instrument cuvette

holder.

• Place the battery in the instrument as described in the “BATTERY REPLACEMENT” chapter.
• Place the instrument on a flat table.
• Do not place the instrument under direct sun light.

To compensate the meter for the sample turbidity or color, the measurement takes place in two
phases. First, the meter is zeroed using the unreacted sample. After the reagents are added the
reacted sample is measured.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.
The blinking “ZERO” indicates that the
instrument needs to be zeroed first.

• Fill a graduated beaker with 50 mL of

sample.

• Add 0.5 mL of HI 93720A-0 Calcium

indicator solution, then swirl to mix.

• In the same beaker add 0.5 mL of

HI 93720B-0

Alkali solution for Calcium,

then swirl to mix.

• Fill two cuvettes with 10 mL of sample

each.

• Add 1 drop of HI 93720C-0 EGTA solution

to one cuvette, replace the cap and swirl
the solution. This is the blank.

• Place the blank into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• Turn the meter on by pressing ON/OFF.

The display briefly shows all tags on.