• Remove the cuvette.

For free chlorine measurement:

add 3

drops of HI 93701A-F (DPD1 indicator) and
3 drops of HI 93701B-F (DPD1 buffer) to an
empty cuvette. Swirl gently to mix, and
immediately add 10 mL of unreacted sample.
Replace the cap and shake gently.

• Immediately replace the cuvette into

the cuvette holder and ensure that the
notch on the cap is positioned securely
into the groove.

• Press READ/

and

the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• At the end of measurement, the instrument

directly displays concentration in mg/L of
free chlorine on the LCD.

For total chlorine measurement,

add 3

drops of HI 93701A-T (DPD1 indicator),
3 drops of HI 93701B-T (DPD1 buffer)
and

1 drop of HI 93701C-T (DPD3

solution) to an empty cuvette. Swirl gently
to mix, and immediately add 10 mL of
unreacted sample.

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.