beautypg.com

Hanna Instruments HI 96731 User Manual

Page 14

background image

14

• Remove the cuvette.

• Add 0.5 mL of HI 93731B-0 Zinc Reagent

B Cyclohexanone. This is the sample.

Note:

To prevent any contamination from

the polycarbonate cap, prior to replacing
it, close the sample cuvette with the
supplied HDPE plastic stopper.

• Replace the cap and swirl the sample for

15 seconds.

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.

Alternatively, wait for 3 minutes and 30
seconds then just press READ/ /TIMER.

In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the instrument

directly displays concentration in mg/L of
zinc on the LCD.

CAUTION:
HI 93731A-0

Zinc Reagent A contains cyanide.

Cyanide, its solutions and hydrogen cyanide
liberated by acids are very poisonous!