beautypg.com

Hanna Instruments HI 96746 User Manual

Page 8

background image

14

15

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• After a few seconds, the display will show

“-0.0-”

. The meter is now zeroed and

ready for measurement.

• Fill one graduated mixing cylinder up to

the 25 mL mark with sample.

• Add the content of one packet of

HI 93746-0

reagent, close the

cylinder and shake well for 30
seconds. This is the sample.

• Fill a cuvette with 10 mL of the sample

up to 1.5 cm below the rim and replace
the cap.

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.

Alternatively, wait for 30 seconds and
just press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the instrument

directly displays concentration in mg/L of
iron on the LCD.

VALIDATION PROCEDURE

Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:

Do not validate the instrument with any standard solutions other than the HANNA CAL

CHECK

TM

Standards, otherwise erroneous results will be obtained. For accurate validation, please

perform test at room temperature: 18 to 25°C (64.5 to 77.0°F)

Turn the meter on by pressing ON/OFF.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.

INTERFERENCES

Interference may be caused by:

• Place the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

Cadmium

above 4.0 mg/L

Chromiun

3+

above 0.25 mg/L

Chromium

6+

above 1.2 mg/L

Copper

above 0.6 mg/L

Manganese above 50.0 mg/L
Molybdenum above 4.0 mg/L
Nitrite ion above 0.8 mg/L
Cobalt

above 0.05 mg/L

Cyanide

above 2.8 mg/L

Mercury

above 0.4 mg/L

Nickel

above 1.0 mg/L

Sample pH should be between 3 and 4 to avoid developed color to fade or turbidity.