13

8 seconds

MAKING A STANDARD %BRIX SOLUTION

MAKING A STANDARD %BRIX SOLUTION

MAKING A STANDARD %BRIX SOLUTION

MAKING A STANDARD %BRIX SOLUTION

MAKING A STANDARD %BRIX SOLUTION

To make a Brix Solution, follow the procedure below:
• Place container (such as a glass vial or dropper bottle that has a cover) on an analytical balance.

• Tare the balance.
• To make an X BRIX solution weigh out X grams of high purity Sucrose (CAS #: 57-50-1)

directly into the container.

• Add distilled or deionized water to the container so the total weight of the solution is 100 g.
• Note: Solutions above 60 %Brix need to be vigorously stirred or shaken and heated in a water

bath. Remove solution from bath when sucrose has dissolved. The total quantity can be
scaled proportionally for smaller containers but accuracy may be sacrificed.

Example with 25 %Brix:
%Brix

%Brix

%Brix

%Brix

%Brix

g Sucrose

g Sucrose

g Sucrose

g Sucrose

g Sucrose

g Water

g Water

g Water

g Water

g Water

g T

g T

g T

g T

g Total

otal

otal

otal

otal

25

25.000

75.000

100.000

TO CHANGE POTEN

TO CHANGE POTEN

TO CHANGE POTEN

TO CHANGE POTEN

TO CHANGE POTENTTTTTIAL ALCOHOL CONVERSION FACTOR (HI 96813

IAL ALCOHOL CONVERSION FACTOR (HI 96813

IAL ALCOHOL CONVERSION FACTOR (HI 96813

IAL ALCOHOL CONVERSION FACTOR (HI 96813

IAL ALCOHOL CONVERSION FACTOR (HI 96813 only

only

only

only

only)))))

To change the potential alcohol conversion factor, follow this procedure.
1. Press and hold ON/OFF key continuously for approximately 8 seconds. The LCD will display

all segments screen followed by a screen with the model number on the primary display and

the version number on the secondary display. Continue pressing the ON/OFF key.

2. While continuing to hold ON/OFF, press the RANGE key. The LCD will show the current

conversion factor on the primary display and “P.ALc” on the secondary display. Continue

holding the ON/OFF key. Press the ZERO key to increase this number. The number will

continually increase until “0.70” is reached then wrap back to “C1”. The user selectable

conversion range is 0.50 to 0.70. C1 stands for “curve 1” (see page 6). When you reach the

conversion factor you desire release the ON/OFF key. The new conversion factor will be

used.