Hanna Instruments HI 98701 User Manual
Page 11
21
20
TO STOP PRINTING
In order to stop the recording mode, press
ALT and LOG keys (press ON/OFF first, if
meter is in sleep mode).
A last printout reporting the number of printed
samples (e.g. S.00009) will be produced.
Notes:
• It is recommended to use the adapter during printing mode,
especially when many printouts are going to be taken.
• Before proceeding with printing, make sure there is enough paper
for your measurements. When the paper is finished the meter will
not advise the operator and the printouts could be lost.
• It is possible to insert a new paper roll during printing session.
• Once in the timed printing mode, the interval cannot be changed.
Exit the timed printing mode first (pressing the ALT and LOG keys)
to set a new interval.
• If the LOG key is pressed while in timed printing mode, a printout
is produced without affecting the running sample number.
LOW BATTERY CONDITION
Printing is automatically interrupted when
battery charge weakens. The last message
"Stop log" will be printed and the LOBAT
symbol will blink on LCD.
Note: When an external adapter or new batteries are connected, a
new printing session must be manually restarted.
LOG ON DEMAND
In measuring or timed printing mode, press LOG to
print the current readings. The printout provides the
following information:
a - Date
b - Sample number
c - Time
d - Readings ("----" means out of
range or probe not connected)
Note: If the user attempts to print while in low battery condition the
message "bAtt" will appear for a few seconds on the LCD.
Note: It is possible to reset the sample number of the print on
demand to 001 (see "TO RESET PRINTING SAMPLE NUMBER"
in "SETUP MODE" paragraph).
HI 98701 - HI 98704
PRINTING FUNCTIONS
Two different modes to print data are available:
1. Timed printing; samples are printed at fixed time intervals. Data
lots from 01 to 16 are available.
2. Print on demand; samples are printed when the LOG key is
pressed. It's possible to perform the print on demand either in
normal mode or in Timed printing mode.
Note: In timed printing mode, channels with no probe connected are
not printed. If all the channels are disconnected, the message
"no connected probe" will scroll across the LCD.
Note: cold junction temperature is not printed.
TIMED PRINTING MODE
To start Timed printing, press ALT and LOG.
The lot number will be displayed for a few
seconds then the LOG symbol will appear on
LCD and a first set of data will be printed.
The printout provides the following informa-
tion:
a - Lot number
b - Logging interval
c - Date (only for the first printed
sample of the lot or of the day)
d - Sample number
e - Time
f - Readings ("----" means out of range).
Note: If no keys are pressed, the meter enters standby mode to
prolong the battery life and only the "LOG" indication will be
visible on LCD. While timed printing during the sleep mode, the
last printed reading will appear briefly on the LCD. To reacti-
vate the LCD press ON/OFF.
TO RETRIEVE LOGGED DATA
Logged data can be viewed on LCD or printed. To view or print logged
data see "TO SCAN LOGGED DATA" in the "SETUP MODE" paragraph.
The logging meters also allow the downloading of logged data to PC.
To download data to PC see "DATA TRANSFER TO PC" paragraph.