Pump calibration procedure – Hanna Instruments HI 84431 User Manual

Page 26

background image

26

Verify: The electrode has been calibrated in 8.30 pH buffer.
• Sample prepar.ation: Add a precisely measured amount of

HI 84431-55 Pump Calibration Solution to a clean beaker

as indicated below:

Fill the beaker up to the 50 mL mark with the distilled or

deionized water. Place the stir bar into the beaker and then place

the beaker into the appropriate place on the instrument top.

• Place the electrode holder on the top of the beaker and secure

it by turning clockwise.

• Immerse the calibrated pH and the temperature electrodes

approximatively 2 cm (0.8”) into the sample to be tested

paying attention not to touch the stir bar.

• Verify if the titrant correspondes to selected titration range.

For Low Range (Strong LR, Total LR) use the HI 84431-50.
For High Range (Strong HR, Total HR) use the HI 84431-51.

After the 5 minutes purging interval has elapsed the “Completed” message is displayed.
Another purge period can be initiated by pressing Restart or press ESC to return to main titrator
screen.

PUMP CALIBRATION PROCEDURE

SETUP alkalinity type and purge unit with titrant before performing the pump calibration.
• Select the corresponding range according to the table below.

The pump calibration must be performed each time the alkalinity type, pump tube, titrant bottle

or the pH electrode is changed. It is recommended the pump calibration be performed before

each set of titrations.

T

I

N

U

e

g

n

a

R

w

o

L

)

e

l

p

m

a

s

L

m

0

5

(

e

g

n

a

R

h

g

i

H

)

e

l

p

m

a

s

L

m

0

5

(

L

/

g

m

0

.

0

0

5

o

t

0

.

0

1

0

0

0

4

o

t

0

0

4

L

/

q

e

m

0

.

0

1

o

t

2

.

0

0

8

o

t

8

Low Range (Strong LR, Total LR) - 2 ml
High Range (Strong HR, Total HR) -20 ml