SPECIFICATIONS

* 1 FTU = 1 NTU

Range

0.00 to 50.00 FTU *

50 to 1000 FTU *

Resolution

0.01 and 1 FTU *

Accuracy

±0.5 FTU* or

±5% of reading (whichever is greater)

Typical EMC Deviation

±2% FS

Calibration

3 point (0, 10 and 500 FTU*)

Light Source

Infrared LED

Light Source Life

Life of the instrument

Light Detector

Silicon Photocell

Battery Type

4 x 1.5V AA alkaline

Battery Life

Approx. 60 hours of use

or 900 measurements

Auto-off

After 5 minutes of non-use

Environment

0 to 50

°

C (32 to 122

°

F);

RH max 95% non-condensing

Dimensions

220 x 82 x 66 mm

(8.7 x 3.2 x 2.6")

Weight

510 g (1.1 lb.)

3

4

5

6

2

1

FUNCTIONAL DESCRIPTION

1) Measurement cell
2) LCD (Liquid Crystal Display)
3) ON/OFF key
4) CAL key, to enter the calibration mode
5) READ/

key, to perform measurements and

to set the date (day and month) of the last
calibration

6) DATE/

key, to display the last calibration

date and to select either the month or the
day of last calibration