Hanna Instruments HI 4002 User Manual

Page 12

background image

12

Procedure
1) Follow sections VIII and IX to prepare sensors for

measurement.

2) Follow section VI to prepare standards/ solution. Stan-

dards should bracket and fall within the range of
interest.
Two mL HI 4000-00 ISA, is added to 100 mL of both
samples and standards Add stir bar and mix before
taking measurements.

3) Follow section VII; General Guidelines to optimize test

set-up.

4) During calibration it is best to start with lower concen-

tration samples first. Wait for a stable measurement
before recording values. Slightly longer equilibra-
tions are required at lower concentrations .

5) To prevent carry over and contamination of samples,

rinse sensors with DIW and dap dry between samples.

Typical Linearity for HI 4002 and HI 4102 Bromide

sensors

-150

-100

-50

0

50

100

150

200

0

1

2

3

4

5

6

7

-Log of Molar Concentration