Procedure:
Follow sections VIII and IX to prepare electrodes for mea-
surement.
1) Follow section VI to prepare standards/ solution. Stan-

dards should bracket and fall within the range of
interest. Standards and solutions should be at the
same temperature.

2) 2 mL of HI 4004-00 is added to 100 mL of both

samples and standards.
Add stir bar and mix before taking measurements.

3) Follow section VII; General Guidelines to optimize test

set-up.

4) During calibration it is best to start with lower concen-

tration samples first. Wait for a stable reading before
reading/ recording values. Permit longer equilibra-
tion times at these levels (3 or 4 minutes).

To prevent carry over and contamination of samples, rinse
sensors with deionized water and remove moisture with
absorbant tissue between samples.

Typical HI 4004 and HI 4104 Linearity

-80

-60

-40

-20

0

20

40

60

80

100

0

1

2

3

4

5

6

-Log of Molar Concentration

mV