VII. General Guidelines

General Guidelines

General Guidelines

General Guidelines

General Guidelines

Concentrated samples (>0.10 M) should be diluted
before measurement. Multiply the final result by the
corresponding dilution factor.

Calibration standards and sample solutions should
have the same ionic strength.

For high ionic strength samples use standard addi-
tion or titration methods.

Calibration standards and sample solutions should
be at same temperature.

The magnetic stirrer may generate heat. Thermally
insulate beaker containing standard or sample from
magnetic stirrer by placing cork or other insulative
sheet between beaker and stirrer plate.

Calibration standards and sample solutions should
be stirred at the same rate using identical sized TFE
coated stir bars.

Rinse electrode pair with distilled or deionized water
between samples and gently dab dry with soft dis-
posable absorbent toweling. Do not rub electrodes.

Avoid Cu(OH)

2

precipitation by maintaining pH be-

tween 4 and 6 pH.

Presoaking the cupric sensor tip in a dilute standard
will optimize response. Use concentrations approxi-
mately10

-3

M.

A scratched, pitted, or oxidized pellet surface can cause
drift, a loss of low level response, or poor repeatability.
Optimum response can be restored by removing the
damaged surface with the microabrasive strip HI 4000-
70. Use gloves to protect skin.

Avoid large changes in temperature (thermal shock)
as it may damage the sensor.