VII. General Guidelines
General Guidelines
General Guidelines
General Guidelines
General Guidelines
•
Concentrated samples (>0.10 M) should be diluted
before measurement. Multiply the final result by the
corresponding dilution factor.
•
Calibration standards and sample solutions should
have the same ionic strength.
•
For high ionic strength samples use standard addi-
tion or titration methods.
•
Calibration standards and sample solutions should
be at same temperature.
•
The magnetic stirrer may generate heat. Thermally
insulate beaker containing standard or sample from
magnetic stirrer by placing cork or other insulative
sheet between beaker and stirrer plate.
•
Calibration standards and sample solutions should
be stirred at the same rate using identical sized TFE
coated stir bars.
•
Rinse electrode pair with distilled or deionized water
between samples and gently dab dry with soft dis-
posable absorbent toweling. Do not rub electrodes.
•
Avoid Cu(OH)
2
precipitation by maintaining pH be-
tween 4 and 6 pH.
•
Presoaking the cupric sensor tip in a dilute standard
will optimize response. Use concentrations approxi-
mately10
-3
M.
•
A scratched, pitted, or oxidized pellet surface can cause
drift, a loss of low level response, or poor repeatability.
Optimum response can be restored by removing the
damaged surface with the microabrasive strip HI 4000-
70. Use gloves to protect skin.
•
Avoid large changes in temperature (thermal shock)
as it may damage the sensor.