beautypg.com

Unit:power – EXFO PM-1100 Optical Power Meter User Manual

Page 4

background image

P/N: 1051923

January 2008

4/4

³

The UNIT:POW command should be named and defined as follows:

³

The Quick Reference Command Tree should be modified as follows to
reflect changes in GPIB commands:

UNIT:POWer

Description This command changes the measurement display units.

Syntax

UNIT:POW

Parameters The parameter can be

³

W: measured value displayed in watts (pw, nw, μw, or mw)

³

DBM: measured value displayed in dBm

³

DB: measured value displayed in dB relative to the current

reference

³

DW: measured value displayed in watts relative to the

current reference

Example

UNIT:POW DBM

Command

Parameter/

Response

Description

MMEM ACQ

DATA

RECA

VALUE? (0 to 1025)

Get acquired

measurement

WAVE? (9999 nm)

Get wavelength

SENS

FREQ CATA?

(99.9;99.9;...)

List sampling rates

POW

WAVE

<9999> [NM]

Set wavelength

WAVE?

(9999 nm)

Get wavelength

UNIT

POW

Set display unit