3.3 Start up

When the EX-5175-EC transmitter is first powered up, it goes through a series of momentary screens, which identify
the instrument model number, serial number and software revision. After all of the momentary screens have been
displayed, the instrument arrives at the Main Gas Display showing the gas concentration and unit of measurement.

Depending on transmitter configuration and calibration condition, the furthest right character in the display may flash a
letter indicating the instrument status. See the Section 4.1.2 below.

3.3.1 Typical Start Up

When power is supplied to the EX-5175-EC, the S/T will display the following sequence of information:

N

OTE

:

Software revision may cause variations of display output.

Example of Display

Function

The instrument: Model EX-5175-EC
Note Toxic is not displayed

The instrument: Serial Number

The instrument: Software Revision

IF the right most character is a flashing W

The instrument is in Warm-up mode

 This should last about 1 minute
 The Signal Output is held at 4mA during warm-up

For

Toxic

Gas

The instrument: Normal Display Mode

Measurement of target Gas

IF the right most
character is a flashing C

The last calibration of the instrument was invalid

 The instrument must be recalibrated

IF the right most
character is a flashing F

There is a sensor fault