beautypg.com

Normal display mode, 5ppm, Exit – ENMET EX-5175 User Manual

Page 14: Zero cal ok, Bad zero pv: 0, Pv: 0 20 cal ok same mv bad sens span, Ma span, Settdsec 0 alarm3, Λ 50, Settdsec 0 alarm1

background image

EX-5175-EC

ENMET Corporation

10


Normal Display Mode

5ppm

M

ENU

H

OLD

the magnet over the

M

ENU

switch for 2 – 4 seconds

to enter the Maintenance Menus

Exit

S

ELECT

Tap the

S

ELECT

switch to return to the Normal

Display Mode.

See Section 5.2.3

M

ENU

Tap the magnet over the

M

ENU

switch to cycle through Maintenance Menus

OR

Zero

Cal OK

If the Zero signal is within Preset Specs the EX-5175-EC will

display Cal OK, See Section 5.2.1

If the Zero signal is not within Preset Specs the EX-5175-EC will

display Bad ZERO

N

OTE

: Some software revisions require the S

ELECT

switch be

tapped to accept the signal.

S

ELECT

M

ENU

Bad ZERO

PV: 0

S

ELECT

Tap the

S

ELECT

switch to initiate Zero adjustment

Apply Calibration Gas until signal value becomes
stable (about 1 to 4 minutes) See Figure 4

If cal is good display will indicate OK or Same

If cal is not within preset “range” display will

indicate Bad Sens

M

ENU

PV: 0

20

Cal OK

Same mV

Bad Sens

Span

S

ELECT

S

ELECT

OR

OR

N

OTE

: You can change the Calibration Gas Level.

H

OLD

the magnet over the

S

ELECT

switch

The

M

ENU

switch changes digit indicated by underscore cursor

The

S

ELECT

switch locks underscored digit and moves to next digit

See Section 5.2.2

S

ELECT

mA Span

M

ENU

S

ELECT

100

To change mA Span set point:
Tap the

M

ENU

switch until mA Span is displayed

Tap the

S

ELECT

switch to display the set point

The

M

ENU

switch changes digit indicated by underscore cursor

The

S

ELECT

switch locks underscored digit and moves to next digit

To return to Normal Gas Display:
Tap

M

ENU

switch until EXIT is displayed

Then tap

S

ELECT

switch

Select

Menu

To change Alarm set points:
Tap Menu switch until Alarm to be changed is displayed
Tap Select switch to display the set point

The

M

ENU

switch: changes digit indicated by underscore

cursor
The

S

ELECT

switch: locks in the underscored digit and

moves to next digit

If change is not within range display returns to first digit
If change is within range display moves to Set Time Delay

Use

M

ENU

and

S

ELECT

switches as above to change time

delay. Between 0 and 5 seconds is allowed

If change is within range display moves to next menu

Λ - Indicates increasing alarm
V - Indicates decreasing alarm

SetTDsec

0

Alarm3

S

ELECT

Λ 50

M

ENU

SetTDsec

0

Alarm1

M

ENU

S

ELECT

Λ 10

SetTDsec

0

Alarm2

M

ENU

S

ELECT

Λ 20

Figure 5: EX-5175-EC Maintenance Menu Flow chart