beautypg.com

ENMET ISA-200-RAL (O) User Manual

Page 20

background image

ENMET Corporation

ISA 200 RAL (O)

16

5.2 Sensor Replacement

5.2.1 CO Sensor

A CO sensor must be replaced when it can no longer be calibrated. To replace a sensor, perform the following steps.

§ Turn off the electrical power. The sample air can continue to flow.

§ Open the display panel and remove the two manifold retention screws and remove the manifold. See Figure 9.

§ Remove the old CO sensor, which is the bottom most sensor, and replace it with a new sensor.

C

AUTION

: New sensors come with a shorting clip that must be removed for proper operation. See Figure 8.

§ Replace the manifold. Observe that the flowmeter reading is correct.

§ Turn on the electrical power.
§ Wait 30 minutes, then recalibrate the sensor per the instructions in Section 5.1.3. As the timer is counting down

during calibration, a voltage is displayed on the screen beside the timing count. This voltage must be between:
.38V – .68V

§ Adjust to: .65V if it is not between .38 – .68V. Adjust this voltage during the last 30 seconds of the calibration

interval, using the bottom most potentiometer on the sensor circuit board. See Figure 10.

§ Replace the manifold housing, and secure the display panel. An instrument without the manifold housing in place

is susceptible to RFI.

5.2.2 Oxygen Sensor

An oxygen sensor must be replaced when it can no longer be calibrated in clean air. To replace the sensor, follow the
general steps given for the replacement of the CO sensor, but effect a replacement of the oxygen sensor, which is the
large sensor in the center position of the manifold. After sensor installation, wait four hours before recalibrating.
When recalibrating, it is not necessary to make a voltage adjustment.

5.3 Humidifier Tube

The humidifier tube is used to assure that the CO and oxygen sensors are not subjected to extremely dry air for a long
period of time, which would decrease their useful life. Change the humidifier tube when changing a CO sensor at the
completion of its useful life.

To change the humidifier tube:

§ Remove the sheet metal cover from the top of the enclosure, exposing the humidifier tube. See Figure 10.

§ Note the coiled arrangement of the tube. Remove the old tube and replace it with the new one.
§ Replace the sheet metal cover.

5.4 Flow Control Orifice

A 0.006-inch diameter orifice is used to set the flow rate and to drop the air pressure. It is located where the air enters
the humidifier tube, see Figure 10. In well-maintained medical air systems, this orifice should not clog. However, if
difficulty is experienced in maintaining flow rate with assured inlet pressure, remove air pressure from the equipment
and examine this orifice; replace it if necessary.

5.5 Particulate Filter Replacement

A filter is installed inside the enclosure, as shown in Figure 10.

To replace this filter:

§ Obtain a new filter

§ Note the correct direction of flow, as denoted by the arrow on the filter body

§ Remove the old filter
§ Replace with the new filter