5.1.2 Automatic Zeroing

A valid key entry sets the instrument in the " run AUTOMATIC ZEROING" position, which enables the setting of the
zero gas concentration point. This is desirable if the zero reference of one of the sensors has drifted over time. For
CO, the zero reference point is 0000 ppm CO; the zero reference point for oxygen is 20.9% oxygen by volume. Note
that the calibration procedure described in section 5.1.3 also includes setting the zero point. If a full calibration is
required, instead of setting just the zero point, press the OPTION switch once; " run FULL CALIBRATION" is
displayed. See section 5.1.3.

T

O SET THE ZERO POINT WITHOUT PERFORMING FULL CALIBRATION

:

§ From the "run AUTOMATIC ZEROING" position, see flow chart.
§ Press the SELECT button; "ATTACH ZERO GAS" is displayed.

N

OTE

: While it is possible to zero the sensor(s) using a sample from a "clean" air line, this entails using the sample as

a standard, and is best avoided. The best zero gas is a cylinder of 20.9% oxygen in nitrogen with no CO
present. This is available in the calibration kit listed in section 6.0

Pressing the OPTION switch at this point aborts the procedure and transfers the display to the " exit MAINTENANCE
MODE" position.

From Valid Key Entry

Example: Automatic Zeroing Flow Chart

T

O CONTINUE THE PROCEDURE

: attach a cylinder of zero gas to the Calibration Port using the calibration adapter, as

shown in Figure 7. Open the cylinder valve, set the regulator at 55 psig, and turn the red handle of the sample-
calibrate valve up toward the calibration port. Let the gas flow for about a minute

ALARM

ACKNOWLEDGE

AUDIO
DEFEAT

FAULT

POWER

SELECT

OPTION

Figure 7: Connection of Calibration Gas Cylinder

Cylinder of Gas

Cylinder Valve
and Regulator

Calibration Port

Sample/Calibration
Valve – Up to Calibrate

O

To Full Calibration

Valid

Invalid

O

S

S

ATTACH

ZERO GAS

run

AUTOMATIC

ZEROING

ZERO=30 .00V

20.9 130 00.0

Exit Maintenance Mode

Return to Operational Display

Return to Operational Display

with

“Fault 01” in place of the PPM reading

O = Option Switch
S = Select Switch