beautypg.com

ENMET ISA-200-RAL (O) User Manual

Page 11

background image

ISA 200 RAL (O)

ENMET Corporation

7

In addition, there is a fault relay, which changes state whenever the instrument is in a fault condition.

The contact positions are given in Table 2:

Table 2: Fault Relay Contacts

Position

Function

Contact

13

Fault

NC

14

Fault

C

15

Fault

NO

The coil of this relay is energized when the instrument is in the non-fault state; the contact conditions given above are
for the non-energized state, which is identical to the fault state.

These relay contacts can be used to operate auxiliary alarms or other functions. Punch a hole at the bottom of the left
side of the enclosure for a wire exit, and use appropriate cable and fittings to preserve the NEMA-12 rating of the
enclosure.

3.4.2 Optional 4-20mA Outputs

Isolated 4-20 mA outputs are available for data logging or other purposes. An output is supplied for each sensor
supplied in a particular instrument, and can be added when a sensor is added in the field. When all three sensors are
supplied, these outputs are available on the terminal strip in the positions given in Table 3:

Table 3: Outputs for 4-20mA

Position

Channel

Function

Range

16

CO

Ground

4 mA = 0 ppm

17

CO

+ 4 to 20 mA

20 mA = 100 ppm

18

Not Used

19

Not Used

20

O2

Ground

4 mA = 0%

21

O2

+ 4 to 20 mA

20 mA = 25.5%

When both sensors are supplied, the sensor for CO output is on positions 16 and 17. The Oxygen output is on
positions 20 and 21. Positions 18 & 19 are not used.

Wiring requirements are the same as for the relays.

3.5 Initial Calibration

If a calibration kit is available, calibrate the CO and O

2

channels of the instrument 24 hours after installation. See

Section 5.0, Maintenance, for calibration instructions. After calibration, be sure to return the red sample-calibrate
valve handle to the down position, pointing toward the sample input port.