CP-60
ENMET Corporation
16
5.2.7 Set New Password
To change the password, you must enter the maintenance menu. Press the
OPTION
switch until “Enter Maint Menu” is displayed
then press
SELECT
switch for the Enter Password menu. Enter the valid password as described in Section 5.2.1.
To set a new password, after inserting a valid password,
Press the
OPTION
switch until; "
Set New Password”
is displayed.
Press the
SELECT
switch; "
Password
❚
1270
" is displayed, with the underscore cursor under the left digit.
Use the
OPTION
switch to change the left digit, when the desired digit is displayed.
Press the
SELECT
switch to lock the digit in place and move the cursor one digit to the right.
When all four digits of the new password have been selected, "
Set New Password
" is displayed.
Record the new password; without it, the maintenance menu cannot be reentered once you exit the Maintenance Menu. If the
password is lost, call
ENMET
customer service personnel.
From the "
Password XXXX
" position,
Press the
SELECT
switch to return to Set New Password section.
Press the
OPTION
switch; to continue to "
exit MAINTENANCE Menu
"
Example: Set Password Flow Chart
5.2.8 Exit Maintenance Menu
From the "
exit MAINTENANCE Menu
" position
Press the
SELECT
switch to resume the operational display.
Press the
OPTION
switch to reenter the maintenance menu at the "
Calibration
" position.
Example: Exit Maintenance Menu Flow Chart
5.3 Channel Activation/Deactivation
Each channel of the
CP-60
can be deactivated and activated as needed.
C
AUTION
:
Deactivated channels will not display or respond to connected sensor transmitters.
A fault indication occurs when a transmitter is not connected to an active channel of the
CP-60
.
To Activate or deactivate a channel of the
CP-60:
Enter the maintenance menu as described in Section 5.2 and press the
OPTION
switch until Exit maint menu appears.
Press and Hold the
OPTION
switch until “
ADV maint menu”
is displayed.
Press the
SELECT
switch to locate the channel to be changed.
Press the
OPTION
switch to change the status of that channel.
Press the
SELECT
switch to accept the change in status. “
ADV maint menu Exit”
is displayed.
Press the
SELECT
switch to return to the operational menu.
O
S
MAINTENANCE MENU
Set New Password
Password
❚0000
O
S
Changes digit indicated by underscore cursor
Locks underscored digit and moves cursor to next digit
O = Press Option
S = Press Select
S
O
MAINTENANCE MENU
Exit maint menu
MAINTENANCE MENU
Calibration
CH-1
CH-3
CH-2
Note:
When in maintenance mode the “_” indicates the location of the cursor, the “
■” block indicates a non-used location.