The alarm level settings for the instrument may be viewed via the menu button (Section 6.5). These values represent the alarm set-
points which have been programmed into the instrument during calibration.
From the menu, the user is prompted to press the up arrow key to view the readings: -






Sequential operation of the up arrow key shows in turn the alarm set points for each sensor. A typical display sequence for a 4-gas
instrument would be:-




Alarm settings for the Oxygen sensor - two settings


or FLAM 1.00 %VOL if in the volume mode

Alarm setting for the Flammable sensor - one setting






Alarm settings for toxic sensor 1, Hydrogen Sulphide - three settings






Alarm settings for toxic sensor 2, Carbon Monoxide - three settings
Important: - The Instantaneous Alarm settings must be greater than both the Short Term and Time-Weighted-Average/Long
Term exposure alarm settings. The Flammable sensor alarm setting cannot be set above 60% LEL.

ALARM LEVELS
PRESS [ ] KEY

INSTANT ALARM
O

2

HIGH 23%

INSTANT ALARM
O

2

LOW 19%

INSTANT ALARM
FLAM 20 % LEL

INSTANT ALARM
H

2

S 15 ppm

STEL ALARM
H

2

S 10 ppm

TWA ALARM
H

2

S 5 ppm

INSTANT ALARM
CO 400 ppm

TWA ALARM
CO 50 ppm

STEL ALARM
CO 200 ppm