Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe

8

EXAMPLE 8. Sample Program using P86 and P22 Instructions

45: Do (P86)
1: 41

Set Port 1 High


46: Excitation with Delay (P22)
1: 1

Ex

Channel

2: 20

Delay W/Ex (units = 0.01 sec)

3: 0

Delay After Ex (units = 0.01 sec)

4: 0

mV

Excitation


47: Do (P86)
1: 51

Set Port 1 Low


48: End (P95)


In the CR200 series dataloggers, use the Portset instruction.

EXAMPLE 9. Sample Program using Portset

Portset (1,1)
Delay(500,msec)
Portset(1,0)


In the CR1000, CR800, CR850, and CR3000 dataloggers, use the Portset
instruction.

EXAMPLE 10. Sample Program using Portset

Portset (1,1)
Delay (1,500,msec)
Portset (1,0)


The above example is not as power efficient as possible and would require AC
power to maintain a sufficient battery charge. If it is necessary to operate an
agitator without AC power available, write the program so that the agitator is
only operated for a short period of time just before the measurement is to be
taken.