Accu-Chek Aviva Expert User Manual
Page 78

70
j
The label on your test
strip container shows the
acceptable ranges for both
the Level 1 and the Level 2
control solutions. The test
result you get should be inside
this range. Ensure you select
and compare the test result to
the correct level of control. For
the level selected, the meter
displays whether the result
is within or outside of the
acceptable range.
Example
Range (mg/dL)
Level 1 25-55
Level 2 255-345
j
When the control test result
is inside the range on the test
strip container, the test strips
and the meter are working
properly.
If the control test result is within
the acceptable range, “Within
range” appears on the detailed
Control Result screen.
Example Control Results
52123_07006322001.indb 70
10/18/13 2:31 PM