YSI 85 User Manual
Page 38
Troubleshooting
Section
12
YSI, Incorporated
Model 85
34
SYMPTOM
POSSIBLE
CAUSE
ACTION
12. Main display reads “LErr”
A. In temperature compensated
conductivity mode, temperature
exceeds the values computed using
user defined temperature coefficient
and/or reference temperature.
B. In cell constant cal mode, temperature
exceeds the values computed using
user defined temperature coefficient
and/or reference temperature.
A. & B. Adjust user defined tempco or
reference temperature. (pg. 10)
13. Main display reads “Err”
(Secondary display reads “ra”)
A. System has failed its RAM test check
procedure.
A. Turn instrument OFF and back ON
again.
B. Return the system for service (pg. 26)
14. Main display reads “Err”
(Secondary display reads “ro”)
A. System has failed its ROM test check
procedure.
A. Turn instrument OFF and back ON
again.
B. Return the system for service (pg. 26)
15. Secondary display reads “rEr”
A. Temperature jumper is set to °F and
reading is >199.9°F but <203°F.
A. Return the system for service. (pg. 26)
16. Main display reads “FAIL”
(Secondary display reads “eep”)
A. EEPROM has failed to respond in
time.
A. Return the system for service. (pg. 26)
17. Readings on main display don’t
change
A. Meter is in recall mode.
A. Press MODE button to return to Normal
Operation (pg. 12)