Read commands – VACUUBRAND VNC 1 User Manual

Page 27

background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. Manual-no.: 999118 / 16/12/2009

page 27 of 35

Read commands

Function

Command Response

Description

Error status

IN_ERR

XXXX

1: Last command to interface incorrect
1: Failure at pressure transducer
1: Overpressure
1: See ERR_1

Error status

IN_ERR_1

XXXX

1: Temperature error
1: Error isolation valve
1: Error cooling water valve
1: External errorr (e.g. at VARIO)

Preset pressure

IN_SP_1

XXXX mbar or

Unit according to preselection

XXXX Torr or
XXXX hPa

Status of

IN_STAT

XXXX

process control

VACUU•LAN

0: Off
1: Process
2: Pressure3: Pressure check

Continuous

0: Off

pumping

1: On

Vacuum control

0: Off
1: Pressure>upper pressure value
2: Preset pressure

upper pressure value

3: Pressure4: End

RC 5 / RC 6

0: Off
1: Warmup
2: Pressrue>preset pressure
3: Pressure

Mode

0: VACUU•LAN
1: Continuous pumping
2: Vacuum control
4: RC 5 / RC 6 management
0: not used

Cooling water

0: Closed

valve

1: Open