Read commands
Function
Command Response
Description
Error status
IN_ERR
XXXX
1: Last command to interface incorrect
1: Failure at pressure transducer
1: Overpressure
1: See ERR_1
Error status
IN_ERR_1
XXXX
1: Temperature error
1: Error isolation valve
1: Error cooling water valve
1: External errorr (e.g. at VARIO)
Preset pressure
IN_SP_1
XXXX mbar or
Unit according to preselection
XXXX Torr or
XXXX hPa
Status of
IN_STAT
XXXX
process control
VACUU•LAN
0: Off
1: Process
2: Pressure
Continuous
0: Off
pumping
1: On
Vacuum control
0: Off upper pressure value 3: Pressure RC 5 / RC 6 0: Off Mode 0: VACUU•LAN Cooling water 0: Closed valve 1: Open
1: Pressure>upper pressure value
2: Preset pressure
1: Warmup
2: Pressrue>preset pressure
3: Pressure
1: Continuous pumping
2: Vacuum control
4: RC 5 / RC 6 management
0: not used