Read commands
Function
Command Response
Description
Actual pressure
IN_PV_1
XXXX mbar or
Unit according to preselection
XXXX Torr or
XXXX hPa
Actual pumping speed
IN_PV_2
XX.X Hz
Time
IN_PV_3
XX:XX h:m
Time for shut down or time of warmup
(only RC 5 / RC 6)
Increase of pressure
IN_PV_4
xxx/min
Only VACUU•LAN
Time of process
IN_PV_5
XX:XX h:m
Only VACUU•LAN
Preselected
IN_CFG
XXXXX
at controller
0: Remote mode off
1: Remote mode on
0: not used
0: not used
0: not used
0: not used
0: No cooling water valve
1: Cooling water valve
0: VACUU•LAN
1: Continuous pumping
2: Vacuum control
4: RC 5 / RC 6 management
Preselected
IN_CFG_1
XXXXXXXX
at controller
0: Remote mode off
1: Remote mode on
0: mbar
1: Torr
2: hPa
0: External error off
1: External error on (evaluation)
2: Stop (control is stopped)
0: Acoustic signal off
1: Acoustic signal on
0: Isolation valve off
1: Isolation valve on
0: Cooling water valve off
1: Cooling water valve on
Control with 0: Valve
1: Pump
2: VARIO
0: VACUU•LAN
1: Continuous pumping
2: Vacuum control
4: RC 5 / RC 6 management