Replacing the fuse – VACUUBRAND MZ 2S NT User Manual
Page 81
+
The pigtail fuses are
integrated into wires
((1), black and blue)
inside the terminal
box. To replace the
fuses it is neces-
sary to replace both
wires completely
(fixed with flat pin
bushings (2)).
Replacing the fuse
➨
Switch off the pump.
➨
Disconnect the electrical power cord before open-
ing the terminal box. After disconnecting from power,
wait five seconds to allow the capacitors to discharge.
+
The replacing of the fuse has to be carried out by a
trained electrician.
+
After replacing the fuse, the pump must be checked for
electric safety (see below)! Identify and eliminate the
cause of failure before switching on the pump again.
1
(1)
(2)
(2)
2
T20
4x
636542